Interlock List
4 Copyright © 2013 EVG
1.1 Leak Sensor
Below the cooling station and the bond chamber there is a tray with a
liquid sensor. If liquid reaches the sensor the system shuts down.
Sensor:
Liquid level switch SK1-8-M12-P-nb
Figure 3 - Leak Sensor
Function:
If liquid reaches the sensor only the water supply stops.
Recovery Procedure:
1) It is necessary to fix the leak.
2) Restart the whole system.
3) Perform an automatic clean up (if available).
4) Start a new process.
Location:
Figure 4 - Leak Sensor