Preventive Maintenance Manual
108 Copyright © 2012 EVG
5.12.2 Close Cover
Check if the status of “Cover” (1) changes from “open” to “closed” when
the cover is closed.
Figure 117 - Status "Cover Closed"
5.13 Check of Sensors
5.13.1 Leakage Sensor
Put a paper strip underneath the leakage sensor to check the
functionality. The sensor must release LED on the sensor will be off,
error message “MACH: Leakage detected” will appear in the SW and
the water flow will be stopped.
If the sensor doesn’t release adjust the sensitivity with the
potentiometer.
1